XI - SERVICING AND
STERILIZATION

11.1 SERVICING THE DEVICE

Important:
The device must be OFF during cleaning/disinfection
procedures.
Do not use abrasive product, sprays or liquids to
clean and disinfect the console.
Avoid the use of cleaning and disinfection products
containing Á ammable agents.
If you do use such a product, make sure that all
the product has evaporated and that no Á ammable
agents remain on the device and its accessories
before using the device again.
The device, the bracket and the multifunction
pedal are not suitable for sterilization.
The console, the control pedal and the bracket shall
be cleaned and disinfected (using alcohol, disinfection
products, dentist disinfecting wipes, S
systematically after each use.
The console must be checked regularly to detect any
problems.
It is important that you check the cleanliness of the
ventilation inlets on the console to prevent abnormal
heating.

11.2 SERVICING THE ULTRASOUND HANDPIECE

Important:
The handpiece is not delivered sterile.
Refer to the user's manual delivered with the
equipment.

11.3 SERVICING TIPS

Important:
Tips are not delivered sterile.
Refer to the user's manual delivered with the
equipment.
12

11.4 SERVICING IRRIGATION LINES

Sterile irrigation line
The disposable irrigation lines supplied by SATELEC
shall be systematically disposed off after each use.
The reuse of a disposable irrigation line may result
in contamination of your patients for which you are
liable.
Do not attempt to resterilize disposable irrigation
lines. Do not attempt to modify irrigation lines.
Sterilizable irrigation line
Refer to the instructions delivered with the
equipment.
To service, clean, decontaminate and sterilize the
sterilizable irrigation line, refer to the speciÀ c
technical maintenance guide.
XII - CHECKING / MAINTENANCE
™ wipes)
EPTOL
In the event of an anomaly, we recommend you
contact the supplier of your device. Do not contact
any other repairer who may render your device
hazardous for you and your patients.

12.1 CHECKING

Check the device and its accessories regularly to
detect any insulation fault or any damage. Replace
if necessary.
It is important to check the cleanliness of the
ventilation grilles on the console ventilation to
prevent abnormal overheating.

12.2 MAINTENANCE

The device does not require any preventive
maintenance other than the checking of accessories
and frequent cleaning/disinfection.
Important: