Channel fault during the Bumptest.
For all sensors "Fail !"
and extra lines "Error: CO2 !"
"Error: CH4 !"
Etc.
Dräger X-am 7000
X - am 7000
ARXX0000
8317400
Pers - ID:
PERSONAL - ID
04 . 01 . 2007
16 : 27
Bumptest
CO2
:
Fail !
Bumptest
CH4
:
Fail !
:
Fail !
Bumptest
O2
Bumptest
CO
:
Fail !
Bumptest
H2S
:
Fail !
Mode:
auto
Error:
CO2 !
Error:
CH4 !
Error:
O2 !
Error:
CO !
Error:
H2S !
11 : 49
Battery:
OK
10.4

Replacing the Sensors

WARNING
!
The device may either be equipped with a DrägerSensor XS Cl
or with a DrägerSensor XS CO
sensors is not permitted. Also, the device may not be equipped
with two sensors of the type XS Cl
XS CO
. An XS Cl
or XS CO
2
2
all other gas sensors. If these requirements are not complied
with, the device will lose its explosion protection.
Switch off the instrument.
1
Remove the sensor grille from
the instrument (secured with
two screws).
Remove the sealing plate.
2
With the aid of the sensor grille
3
remove the sealing plate,
taking care not to damage the
sensor membrane!
Maintenance
. A combination of these two
2
or two sensors of the type
2
sensor may be combined with
2
1
01323727_1.eps
2
3
01923727_1.eps
83