Maintenance
or
2. Automatic start in the BumpTest Station (order no. 83 18 909).
Prerequisite:
-
The instrument must be configured to start the Bumptest automatically
via the Dräger CC Vision software (version 5.3.3 or higher).
-
The instrument must be inside the BumpTest Station with the
Mobile Printer connected and switched on (order no. 83 19 310).
Sequence:
Supply test gas, see the Instructions for Use for the BumpTest Station.
Position an activated Mobile Printer in front of the IR interface of
the BumpTest Station.
Open the flap of the BumpTest Station and insert the instrument.
-
The automatic Bumptest procedure is activated.
-
The Bumptest is successful if the A1 alarm is triggered in the
sensors involved and the entered Bumptest concentration has
been reached.
Printing the result of the Bumptest with the Dräger Mobile Printer
Prerequisite:
-
The instrument must be configured to print the result of
theBumptest with the Dräger Mobile Printer via the Dräger CC
Vision software (version 5.3.3 or higher).
Printing the result
-
If Bumptest was started manually:
after the displayed result of the Bumptest has been confirmed by
pressing the »
« key.
M
-
If Bumptest was started automatically:
automatically, if Bumptest was successful.
The printout contains the following information:
82
Examples
Bumptest passed.
Variable specifications:
1
*
:
"auto" or "manu"
2
*
:
can be configured by
the user
3
*
:
"OK 11:49" or "ALARM"
Instrument
fault
during
the
Bumptest.
For all sensors "Fail !"
and extra line "Error: Device !"
X - am 7000
ARXX0000
8317400
2
Pers - ID:
PERSONAL - ID
*
04 . 01 . 2007
14 : 57
Bumptest
CO2
:
Pass
Pass
Bumptest
CH4
:
Bumptest
O2
:
Pass
Bumptest
CO
:
Pass
Bumptest
H2S
:
Pass
1
Mode:
auto
*
Battery:
OK
11 : 49
*
3
X - am 7000
ARXX0000
8317400
Pers - ID:
PERSONAL - ID
04 . 01 . 2007
15 : 26
Bumptest
CO2
:
Fail !
Bumptest
CH4
:
Fail !
Bumptest
O2
:
Fail !
Bumptest
CO
:
Fail !
Fail !
Bumptest
H2S
:
Mode:
auto
Error:
Device !
Battery:
OK
11 : 49