The Menu » Calibration «
-
The corresponding error message is displayed.
To return to the menu:
Press the »
« key
OK
To terminate fresh-air calibration:
Select » Menu « and press the »
6.3

Mixed-gas calibration

-
This function is used when the sensitivity of several sensors is
to be calibrated in a single step.
-
All activated sensors must be warmed up (no warning icon »
on the display.
NOTICE
i
i
Observe the information about cross-sensitivity of certain
gases shown in the Data Sheet for the sensor(s) to be
calibrated.
Select » Calibration « and press the »
Select » Mixedgas cal. « and press the »
-
The values for the mixed gas are displayed, for example:
Supply mixed gas
CO
:
2.5 vol. %
2
CO
:
250 ppm
-
The calibration gas, the concentration and the units are the same as
for the calibration of the sensitivity as described on page page 58.
-
If desired, the calibration gas concentration can be changed as
follows:
Use the »
« key to select the appropriate line and press
M
the »
« key.
OK
56
« key.
OK
«
« key.
OK
« key.
OK
Set the calibration-gas concentration with the »
keys and confirm the setting with the »
Apply the mixed gas (see the information in the section
"Maintenance" page 78) to the sensor.
Select » Calibration « and press the »
When the displayed measured values have settled:
Select » Confirm « and press the »
-
The message » Remove Gas ! « is displayed
Turn off the flow of mixed gas through the sensor.
An error is indicated in the case of any fault, such as a sensor or
instrument fault, if the sensor is not yet warmed up or if the data and
time have not been set. If this happens:
Use the »
« key to select the appropriate line and press
M
the »
« key.
OK
-
The corresponding error message is displayed.
To return to the menu:
Press the »
« key
OK
To terminate mixed-gas calibration:
Select » Menu « and press the »
« and »
«
M
« key.
OK
« key.
OK
« key.
OK
« key.
OK