Adjustable range:
Measured gas
see the appropriate sensor data sheet
Short gas name
4 characters
Long gas name
10 characters
Full-scale deflection
see the appropriate sensor data sheet
Response factor
0.1 to 20.0
Molecular weight
2 to 250 g/mol
A2
greater than 0, less than full-scale deflection
A1
greater than 0, less than A2
Determination of the response factor of a customer-specific gas
Configure PID sensor to isobutylene (IBUT).
Adjust zero point and sensitivity.
Apply gas to the sensor using customer gas of known concentration
(e.g. 150 ppm) and wait for stable measurement value.
Make a note of the displayed measurement value (e.g. 120 ppm)
and determine the response factor as follows.
RF = (desired value / actual value) = (150 ppm / 120 ppm) = 1.25
Turn off the gas flow and set the sensor to the freely configurable
gas (VOC1, VOC2,...).
After the short and the long gas names have been entered, enter the
response factor of the gas to be measured by pressing the »
and »
« keys and confirm the setting by pressing the »
M
Dräger X-am 7000
7.3.5

Adjustable calibration interval

-
Use this function to enter the calibration interval setting of the sensor.
Select the menu items » Settings \ Sensors \ CA-CH4 \ Cal.-
Interval « in this order, pressing the »
Set the calibration interval using the »
confirm by pressing the »
-
Only one calibration interval setting is entered for each EC, IR and
PID sensor.
7.3.6

Mean value period

-
This function can be used to set the mean value period for
exposure evaluation.
Select the menu items » Settings \ Sensors \ EC–CO \ Expo-
Param. \ Mean value period « in this order, pressing the »
key after each selection.
Select the desired configuration from the list and, if more than one
configuration has been enabled for activation, press the »
key (»
«), see » Select status « page 69.
Use the »
«
confirm the setting with the »
« key.
OK
Adjustable range: 1 to 60 minutes.
The Menu » Settings «
« key after each selection.
OK
« and »
« key.
OK
« and »
« keys to set the mean value period and
M
« key.
OK
« keys and
M
«
OK
«
OK