Maintenance

Maintenance intervals

The device should be inspected annually by suitably
qualified personnel.
Charge the lead battery after each application, at
the latest however after the battery alarm has been
triggered.
Maintenance by suitably qualified personnel – every
year.
– The inspection intervals must be established in each
individual case and shortened if necessary,
depending on technical safety considerations,
engineering conditions, and the technical
requirements of the equipment.
– We recommend that a service agreement be
concluded with Dräger and that repairs also be
carried out by Dräger.
NOTICE
For the gas detection instruments used, the
maintenance intervals specified in the relevant
Instructions for Use apply.

Replacing the batteries

Explosion hazard!
Do not replace the batteries in areas subject to
explosion hazard! Batteries are part of the Ex approval.
Only the following types may be used:
– Battery pack, small - order no. 83 20 644
– Battery pack, large - order no. 83 20 646
Switch off the device (refer to "Switching off the
device" on page 30).
Unfasten the screws
(M5 cylinder-head
screw with internal
hexagon) on the
bottom side of
the housing.
Lift the upper part
of the housing and
disconnect the plug
connection from the
bottom plate.
WARNING
Release the four M5
nuts.
Disconnect the cable
connections from the
bottom plate.
Replace the old battery
block with a new one.
Re-establish the cable
connection to the
bottom plate.
Check the correct
position of the O-ring.
Tighten the four M5
nuts.
Establish plug connection to the bottom plate.
Place the upper part of the housing on the bottom
part (note preferred position).
Tighten the screws (M5 cylinder-head screw
with internal hexagon) at the housing bottom
(120 Ncm ±20 Ncm).
NOTICE
It is recommended to fully charge the device after
the battery block has been replaced.
Explosion hazard!
Do not throw used batteries into a fire and do not
open them with force.
Dispose of the batteries in accordance with national
regulations.

Charging the batteries

Explosion hazard!
Do not charge underground or in explosion-hazard
areas.
The chargers are not designed in accordance with
the guidelines for firedamp and explosion protection.
Persons wearing implants may only handle the
powered induction charger if their if the pacemakers
and active implants conform to the relevant legal
requirements. Dräger only guarantees conformity with
Directive 2004/108/EC.
NOTICE
The transmitter coil of the induction charger generates
a weak magnetic alternating field. During operation, all
requirements of the relevant standards regarding
electromagnetic faults are observed. The legal
requirements of Directive 2004/108/EC are met.
WARNING
WARNING
CAUTION