The Menu » Settings «
7.3.7

STEL

-
This function can be used to set the alarm threshold for the
exposurealarm STEV.
Select the menu items » Settings \ Sensors \ EC–CO \ Expo-
Param. \ Short time value « in this order, pressing the »
after each selection.
Select the desired configuration from the list and, if more than one
configuration has been enabled for activation, press the »
«), see » Select status « page 69.
Set the short time value with the »
confirm the setting with the »
Adjustable range: 0 to the entered full-scale deflection.
7.3.8

Short time period

-
This function can be used to set the short time period for exposure
evaluation.
Select the menu items » Settings \ Sensors \ EC–CO \ Expo-
Param. \ Short time value « in this order, pressing the »
after each selection.
Select the desired configuration from the list and, if more than one
configuration has been enabled for activation, press the »
«), see » Select status « page 69.
Use the »
« and »
M
confirm the setting with the »
Adjustable range: 1 to 60 minutes.
68
« and »
« keys and
M
« key.
OK
« keys to set the short time value and
« key.
OK
7.3.9

Total exposure

-
This function can be used to set the alarm threshold for the
exposure alarm TWA.
Select the menu items » Settings \ Sensors \ EC–CO \ Expo-
« key
Param. \ Total exposure « in this order, pressing the »
OK
after each selection.
Select the desired configuration from the list and, if more than one
« key
configuration has been enabled for activation, press the »
OK
«), see » Select status « page 69.
Use the »
in ppm) and confirm the setting with the »
Adjustable range: 0 to the entered full-scale deflection.
7.3.10

Active selection

-
This function can be used to activate one of several (1 to 5) status
records for each sensor.
Select the menu items » Settings \ Sensors \ CA–CH4 \ Cfg.
« key
records \ Active selection « in this order, pressing the »
OK
after each selection.
Select the desired status record from the list and press
« key
the »
« key to activate it (»
OK
OK
« and »
« keys to set the total exposure (value
M
OK
«).
« key
OK
« key
OK
« key.
« key
OK
Dräger X-am 7000