The Menu » Settings «
Select the desired unit from the list and press the »
activate it (»
«).
Select the LEL / Vol.-% conversion factor with the »
»
« keys and press the »
M
Set the full-scale value with the »
confirm the setting with the »
Set the alarm threshold A1 with the »
confirm the setting with the »
Set the alarm threshold A2 with the »
press the »
« key to confirm the setting.
OK
The complete configuration is now displayed and can be
confirmed by pressing the »
Adjustable range:
Measured gas
see the appropriate sensor data sheet
Full-scale deflection
see the appropriate sensor data sheet
LEL Factor
0.1 Vol.-% – 2x default LEL factor of the
appropriate measuring gas
A2
A2 > A1, A2 < full-scale deflection
Exception: for LEL-based units (%UEG, %LEL,
%LIE) A2 has to be  60 %LEL
A1 > 0, A1  A2
A1
-
If the Smart PID sensor is to be used for detecting a customer-
specific gas which is not entered in the sensor data base, gases
provided for this purpose (VOC1, VOC2, ...) can be configured
accordingly.
Select the menu items » Settings\Sensors\PID-IBUT\ Gas cfg. «
in this order, pressing the »
Select VOC1 (or VOC2, VOC3) from the list and press the »
key to activate it (»
«).
66
« key to
OK
« and
« key to confirm the setting
OK
« and »
« keys and
M
« key.
OK
« and »
« keys and
M
« key.
OK
« and »
« keys and
M
« key.
OK
« key after each selection.
OK
Enter the short gas name (max. 4 characters) for the gas to be
measured and press the »
Enter the long gas name (max. 10 characters) for the gas to be
measured and press the »
Enter the gas-specific response factor of the gas to be measured
and press the »
OK
Information on the response factors is available from your local
Dräger Safety branch office. The response factor can also be
determined by the user (see below-mentioned instructions).
CAUTION
!
If an incorrect value is entered, the gas display may also be
incorrect!
Select the desired unit from the list and press the »
activate it (»
«).
If a weight-oriented unit has been selected, enter the molar weight
of the gas to be measured by pressing the »
and press the »
OK
Set the full-scale value with the »
confirm the setting with the »
Set the alarm threshold A2 with the »
press the »
« key to confirm the setting
OK
Set the alarm threshold A1 with the »
confirm the setting with the »
A configuration overview is displayed and can be confirmed by
pressing the »
OK
«
OK
« key to confirm it.
OK
« key to confirm it.
OK
« key to confirm it.
« and »
« key to confirm.
« and »
M
« key.
OK
« and »
« and »
« key.
OK
« key.
Dräger X-am 7000
« key to
OK
« keys
M
« keys and
« keys and
M
« keys and
M