No device function while
operating with CURIS® RF
Generator
No device function while
operating with CURIS®
flow one-pedal footswitch
with push-button
The LED status indicator
(5) lights up red and the
error message "E" is
shown on the display
after self-test sequence
The silicone segment was
pinched during insertion into
the rotary pump
A non-compatible bipolar cord
and tubing set is used.
The rollers located in the rotary
pump (7) must be replaced
CURIS® RF Generator
interconnecting cable is not or
not correctly inserted into the
connection socket (9)
CURIS® RF Generator
interconnecting cable
is defective
Internal defect
CURIS® flow one-pedal
footswitch cable is not or not
correctly inserted into the
connection socket (9)
CURIS® flow one-pedal
footswitch with push-button is
defective
Internal defect
Internal defect
Open the rotary pump lock lever (8)
and make sure the silicone
segment is not pinched while
closing the lock lever
Use only from the manufacturer
recommended bipolar cord and
tubing sets
Replace rotary pump rollers.
Contact your Sutter representative
or the manufacturer
Check the connection of the
CURIS® RF Generator
interconnecting cable
Replace CURIS® RF Generator
interconnecting cable. Contact
your Sutter representative or the
manufacturer
Device is defective, service case.
Contact your Sutter representative
or the manufacturer
Check the connection of the
CURIS® flow one-pedal footswitch
Replace CURIS® flow one-pedal
footswitch with push-button.
Contact your Sutter representative
or the manufacturer
Device is defective, service case.
Contact your Sutter representative
or the manufacturer
Turn the irrigation pump off with the
power switch (1) on the front of the
device, check all connections, and
turn the device back on with the
power switch. If the LED status
indicator (5) still lights up red,
contact your Sutter representative
or the manufacturer.
Device is defective, service case
14