Consider such interference if inexplicable phenomena are observed on the device. Proceed as
follows to restore the proper functioning of the device:
1.
Set up the devices at a safe distance from each other, observe their functionality, and check
them for plausibility.
2.
Make sure that cables do not touch each other, since electromagnetic coupling may occur
during energy input of the HF surgery device.
3.
Take all necessary precautions to avoid malfunctions.
NOTE
Observe the instructions and recommendations of the HF surgery device
manufacturer
Only approved accessories specified by the manufacturer may be used so that the device is not
negatively influenced by electromagnetic phenomena. Furthermore, this ensures that the
electromagnetic interference emission is maintained as determined in the type test.
6 Cleaning and disinfection
For cleaning and disinfection, turn the device off, disconnect it from the mains network, and remove
connected components and accessories. Liquids must not be allowed to get into the interior of the
device when applying cleaning agents and disinfectants.
Submerging or spraying the device can lead to hazards or destroy the irrigation pump.
Use the wiping method for cleaning and disinfection.
Use a cloth moistened with a mild soap solution or 70% isopropanol solution for cleaning. After
cleaning, disinfect the surfaces with a pH-neutral, approved disinfectant on a detergent-alcohol
basis with up to 70% alcohol. Always follow the instructions of the disinfectant manufacturer for
disinfection.
Ensure that cleaning agents and disinfectants are fully removed or evaporated before using the
device.
Visual inspection: The sockets of all connections and the plugs of the connecting cables must not
be soiled in any way.
7 Error display and troubleshooting
If an error in the self-test sequence or during the operation is detected, the LED status
indicator (5) lights up red and the error message "E" is shown on the display. The
irrigation pump then stops operating, and operation cannot be resumed.
For troubleshooting see below table or turn off the irrigation pump with the power switch (1) on the
front of the device, check all connections, and turn the device back on with the power switch.
If the LED status indicator (5) still lights up red, contact your Sutter representative or the
manufacturer.