4.2 Decommissioning

No special measures are required to take the device out of operation.
Turn the device off with the power switch on the front of the device (1).
Subsequently, open the lock lever of the rotary pump (8) and remove the inserted tubing
set. Then close the lock lever of the rotary pump (8) again.

5 Safety measures and precautions

WARNING
To avoid danger for the patient, operating personnel, or third parties, the device must
always be used carefully and strict compliance with the operating and safety instructions
is required!
The irrigation pump is intended for operation exclusively by qualified medical personnel.
Verify the correct functioning of the mains and interconnecting cable before each surgical
use and replace if needed.
Verify the correct flow direction while inserting the tubing set into the rotary pump! Failure
to observe the indicator arrow can cause injury to the patient!
Only use irrigation fluids that met the needs of the medical procedure, and which are
suitable for medical use.
Verify the correct functioning of the CURIS
Always keep the lock lever of the rotary pump closed during operation – do not open it.
Failure to comply can cause injury to operating personnel.
The user must not touch the CURIS
time.
Serious incidents that occur in relation to the product must be reported to the
manufacturer and the responsible public authority in the member state in which the user
and/or patient resides.
Portable and mobile HF communication equipment may affect electromedical devices.
See the guidelines and manufacturer declaration for electromagnetic compatibility in
Chapter 10.2
The CURIS
flow Irrigation pump has no safety measures against burns with the application of HF
®
surgery devices.
The CURIS
flow Irrigation pump can be used in conjunction with HF surgery devices. However,
®
powerful electromagnetic interference produced in the immediate vicinity of electric motors, high-
voltage power lines, PCs, monitors, or other – possible defective – electric appliances may interfere
with the functioning of the device in some cases.
11
flow Irrigation pump before each use.
®
flow Irrigation pump and the patient at the same
®