3.5 System start and self-test

Turning the device on and off:
The device is turned on and off by using the power switch on the front of the device (1).
When it is switched on, an automatic self-test is performed. The display (2) and LED
status indicator (5) light up briefly. When the self-test has successfully completed, the
LED status indicator (5) turns off and the last used flow setting is displayed, indicating
that the irrigation pump is operational.
An error in the self-test sequence indicates a device defect, the LED status indicator (5)
lights up red and the error message "E" is shown on the display. See "Error display and
troubleshooting" in
NOTE
The factory flow setting "0" is shown the first time the device is powered up.

3.6 Inserting the bipolar cord and tubing set

WARNING
Make sure that the CURIS
set. Failure to comply can result in user injury!
NOTE
Sutter recommends using compatible bipolar cord / tubing sets or irrigation tubing.
Using other bipolar cord / tubing sets or irrigation tubing may cause deviating fluid
volumes (excess or short dosing of the irrigation fluid) and failure of the irrigation pump.
The CURIS
tubing and cord set (REF: 6790-100-004, 6790-100-003) manufactured by Stryker
and the sterile CODMAN® Integrated Bipolar Cord and Tubing sets (REF: 9190001RP,
9190002RP).
NOTE
When using a sterile bipolar cord / tubing set or irrigation tubing, verify that the sterile
packaging is intact before use.
Do NOT use the bipolar cord / tubing set or irrigation tubing if:
-
The sterile packaging or product is visibly damaged
-
The packaging has been opened
-
Applied sterility indicators do not have the required color
-
The expiration date has been exceeded.
If CURIS
flow Irrigation pump is used as stand-alone device, the following points 8 and 9 are not
®
applicable or required.
7
Chapter
7.
flow Irrigation pump is turned off while inserting the tubing
®
flow Irrigation pump can be used with the sterile Bipolar irrigator integrated
®