Device fault
There is no device
function and the
elements on the front
remain off
Flow setting control
element (3) has no
function
Device is switched on,
but no irrigation fluid is
conveyed
Not enough irrigation
fluid is conveyed
13
Cause
No power supply
Mains cable is not or not
correctly inserted into the
socket or into the mains cable
connection socket (10)
Device is not switched-on
Internal power supply defective
Rotary encoder is defective
Rotary pump lock lever (8) is
open
The silicone segment was
pinched during insertion into
the rotary pump
The silicone segment was not
inserted correctly into the rotary
pump (7)
A non-compatible bipolar cord
and tubing set is used.
Irrigation fluid bag is empty
The selected flow setting is „0"
Pause function is activated
There is not enough irrigation
fluid
Selected flow setting is too low
Irrigation fluid bag is not
connected correctly with the
spike of the irrigation tubing
Luer connection is not
connected correctly with the
instrument
Corrective action
Check the power supply
Check the connection of the mains
cable
Switch on the device by using the
power switch on the front of the
device (1).
Device is defective, service case.
Contact your Sutter representative
or the manufacturer
Replace rotary encoder.
Contact your Sutter representative
or the manufacturer
Close the rotary pump lock lever
(8)
Open the rotary pump lock lever (8)
and make sure the silicone
segment us not pinched while
closing the lock lever
Open the rotary pump lock lever (8)
and make sure the silicone
segment is inserted according to
the flow direction indicator (7)
Use only from the manufacturer
recommended bipolar cord and
tubing sets
Replace the irrigation fluid bag
Increase the flow setting by turning
the control element (3) clockwise
Deactivate the pause function with
the pause button (6) on the front of
the device
Make sure that sufficient irrigation
fluid is available, if needed change
the irrigation fluid bag
Increase the flow setting by turning
the control element (3) clockwise
Check that the spike of the
irrigation tubing is inserted
correctly
Check that the luer connection of
the irrigation tubing is fully inserted