3.1 Setup for use with the CURIS
Connect either end of the supplied CURIS
(REF: 93008120) to the connection socket (9) on the rear of the irrigation pump.
Connect the other end of the interconnecting cable to the connection socket
marked "PUMP" on the rear of the CURIS
PUMP

3.2 Setup for use with the corresponding footswitch

Connect the CURIS
option (REF: 360115), to the connection socket (9) on the rear of the irrigation
pump.

3.3 Potential equalization connection

Potential equalization is a connection of device housings with good electrical
conductivity. It is intended to ensure that the devices consistently maintain the
same electrical potential, even in case of an electrical defect. For optional
potential equalization, use the potential equalization connection (11). Push the
potential equalization cable on firmly until it snaps into place. The potential
equalization cable is not included in the scope of delivery.

3.4 Mains connection

The device is equipped with a multi-voltage power supply. It can be operated in
the following voltage range without switching:
Connect the mains cable to the connection socket (10) on the rear of the device
and connect the other end of the mains cable to an outlet.
For the all-phase, full disconnection of the device in case of danger, either the
connection socket on the device or the outlet where the mains cable is plugged
in should be left accessible.
No special measures are required to take the device out of operation.
RF Generator
®
flow one-pedal footswitch with push-button, available as an
®
WARNING
To exclude the risk of electric shock, this device may only be connected
to a mains network with protective conductor.
100 – 240 V AC, 50 / 60 Hz
NOTE
Inspect the mains cable and plug for proper function and damage before
each use.
RF Generator interconnecting cable
®
RF Generator.
®