Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity - PNEUMO SURE High Flow Insufflator
The device PNEUMO SURE High Flow Insufflator is intended for use in the electro-
magnetic environment specified below. The user/operator of the insufflator
should make sure the device is operated within such an environment.
Electromagnetic
Test level
interference
immunity tests
Conducted HF
3 Veff
interference
150 KHz to
quantities
80 MHz
according to IEC
3 V/m
61000-4-6
80 MHz to
Radiated HF
2.5 GHz
interference
quantities
according to IEC
61000-4-3
Note 1: The higher frequency range applies for 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
a
The field strength of stationary transmitters such as base stations of wireless
phones and cell phones, ham radio operators, AM and FM radio and TV stations
can theoretically not always determined in advance. A study of the installation
site should be considered to determine the electromagnetic environment con-
cerning the stationary transmitter. If the field strength measured at the usage
Compliance
Electromagnetic environ-
ment guidelines
In compliance
Portable and mobile wire-
less devices should not be
used in closer proximity to
the device (including
cables/lines) than the rec-
ommended safety dis-
tance calculated based on
the transmitting fre-
quency of the applicable
formula. Recommended
safety distance:
d = 1.2P for 150 KHz to
80 MHz;
d = 1.2P for 80 MHz to
800 MHz,
d = 2.3P for 800 MHz to
2.5 GHz
With P as the rated output
of the transmitter in
watts [W] according to
the information provided
by the manufacturer of
the transmitter and d as
recommended safety dis-
tance in meters [m].
The field strength of sta-
tionary transmitters for
all frequencies tested on
a
site
should be lower
than the concordance
b
level.
Interference is possible in
the proximity of devices
featuring the following
pictograph.
Electromagnetic compatibility
EN
85