Electromagnetic compatibility
EN
84
14.5
Guidelines and Manufacturer's Statement – Electromagnetic
Emissions
The device PNEUMO SURE High Flow Insufflator is intended for use in the electro-
magnetic environment specified below. The user/operator of the insufflator
should make sure the device is operated within such an environment.
Emitted interference
Compliance
measurements
HF emission according
Group 1
to CISPR 11
HF emission according
Class B
to CISPR 11
Emission of harmonic
Class B
oscillations according
to IEC 61000-3-2
Emission of voltage
In compliance
fluctuations / flickers
according to IEC 61000-
3-3
Electromagnetic environment guide-
lines
The PNEUMO SURE High Flow Insuf-
flator uses HF energy only for its
internal functions. Therefore, the
device's HF emission is very low and
it is unlikely that devices in close
proximity will experience interfer-
ence.
The PNEUMO SURE High Flow Insuf-
flator is suitable for use in all facili-
ties including those in residential
areas and those directly connected
to a public utility network also sup-
plying buildings used for residential
purposes.