Herunterladen Inhalt Inhalt Diese Seite drucken

Initial Device Startup; Gas Connection - Stryker PneumoSure Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Initial Device Startup

EN
Delivery inspection
Setting up the device
Mains connection
Grounding contact
Only for U.S. operators
Potential equalization
14
4
Initial Device Startup
Always check all parts and accessories of the device immediately after receiving
the shipment. The manufacturer considers only replacement claims that have
been immediately submitted or reported to a sales representative or an autho-
rized service company.
Place the device on a level surface and install in a dry environment. The ambient
temperature and humidity must meet the requirements mentioned in chapter
16 Technical Data, page 89.
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not use in an area where flammable anes-
thetic gases are present.
WARNING!
Position the device such that it is easy to operate and disconnect.
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains connection cable may
be plugged only into a properly installed, grounded safety wall socket (shock-
proof socket) (see DIN VDE 0100-710). Read the device label located in rear of de-
vice (type plate) to determine the operating voltage of the device.
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 or IEC
60320. Grounding will only be reliable if the equipment is connected to a corre-
sponding hospital grade socket.
Integrate the device into the potential equalization system as specified by local
safety rules and regulations.
4.1

Gas connection

WARNING!
Medically pure CO
2
Make sure to use only medically pure CO
mixtures of gases, high pressure compressed gases, gases with entrapped liq-
uids, or polluted gases must not be used with this device.
Use a high-pressure tube to connect a CO2 gas cylinder to the rear gas inlet con-
nection or connect to centralized CO2 gas supply.
. Other gases (i.e., helium, N
2
2
O, argon),

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis