WARNING!
Embolism
Improper placement of the insufflation instrument could cause insufflation of
gas into a vessel, resulting in air or CO
CO
embolism, perform initial insufflation at a low flow rate and ensure that the
2
insufflation instrument is correctly positioned. Check the position of the insuf-
flation instrument immediately if the actual pressure rapidly reaches the nomi-
nal pressure value. CO
embolisms can also be caused by a high intra-abdominal
2
pressure. Avoid high-pressure settings and close damaged blood vessels at once.
WARNING!
Additional insufflation sources
The use of additional insufflation sources increases the intra-abdominal pres-
sure. Continuously monitor intra-abdominal pressure over the course of the en-
tire insufflation if additional sources are used.
WARNING!
Device-inherent dangers
Please read the general risks and dangers information in chapter 2.1 Hazards,
page 5 and chapter 3.3 General Device-Inherent Dangers, page 8.
8.2

Selecting Bariatric Operating Mode

1. Selection of the insufflation mode can be done only while insufflation is
stopped. If Bariatric operating mode is not yet set and displayed, press the
function field Change Mode for 2 seconds to get the insufflation mode over-
view.
2. The display depicts an insufflation modes overview. Select the desired insuf-
flation operating mode by pressing the Bariatric function field. *Inactive op-
erating modes are displayed in gray.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the Configuration
menu (see chapter 10.1 Configuration menu I, page 60).
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
embolisms. To reduce the risk of air or
2
Set Pressure
Actual Pressure
Set Flow
mm Hg
mm Hg
l/min
0
15
3
0.0
Liter
Mode: High Flow
(1)
Change
Actual
START
Mode
0
l/min
(2)