Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

5.3

Touch screen display

(1)
Set Pressure
(15)
mm Hg
15
(14)
(13)
Mode: High Flow
Change
Mode
(12)
(11)
The above depiction of the touch screen also shows all display and function
fields. Field (9) serves as actual flow display (depicted without frame while insuf-
flating) and also as menu function field (depicted with frame).
Field (12) serves as insufflation operating mode display (depicted without frame
while insufflating) and also as control field for selecting the insufflation operat-
ing mode (depicted with frame).
Press the function field (9) or (12) depicted with frame and hold for 2 seconds to
trigger functions or set values. Additional explanations for individual elements
are presented in the subsequent respective control element descriptions.
The status of the gas supply is monitored by the device and indicated with sym-
bols and acoustic signals (see chapter 11 Safety functions for gas pressure display
information).
The following gas bottle pressures are displayed:
> 50 bar
40 - 50 bar
30 - 40 bar
15 - 30 bar; Three warning signals can be heard and the message
"Change gas tank" is displayed. User is advised to obtain a replace-
ment tank.
< 15 bar; Three warning signals can be heard and the message
"Check gas supply" is displayed. Replace gas tank immediately.
If gas supply pressure declines further, there are warnings to remind the user to
replace the gas tank immediately. Five warning signals can be heard and the
message "Check gas supply" is displayed at < 5 bar and again at 0 bar. Insufflation
stops at 0 bar.
(2)
(3)
(4)
(5)
Actual Pressure
Set Flow
mm Hg
l/min
0
3
0.0
Liter
Actual
START
0
l/min
(10)
Fig. 5-3
(1)
(6)
(2)
(3)
(4)
(7)
(5)
(6)
(8)
(7)
(8)
(9)
(10)
(9)
(11)
(12)
(13)
(14)
(15)
Gas supply displays
Gas supply with gas bottle
Operating the Device - General
Screen displays
Continuous pressure reading dis-
play
Gas supply display
Actual pressure display
Gas consumption display/func-
tion field for reset
Gas heating connected/ready
Increasing nominal gas flow
Nominal gas flow display
Decreasing nominal gas flow
Actual gas flow display/ menu
function field
START/STOP function field
Status display/error and warning
messages
Insufflation operating mode dis-
play/selecting insufflation operat-
ing mode
Decreasing nominal pressure
Nominal pressure display
Increasing nominal pressure
EN
17

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis