Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker PneumoSure Gebrauchsanleitung Seite 79

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Safety functions
NOTE!
EN
Please make sure that neither of the two tubes is connected to a trocar or closed
off during a device check. The continuous pressure measurement cannot be acti-
vated otherwise.
If the power fails for less than 1 second, all settings are retained. If the power fails
Power failure
for an extended time period, the device will function as it does when it is being
restarted.
If a device malfunction occurs that prohibits any further use of the device or di-
minished its safety, the display depicts only
Text message - >Call service
The following text messages may be displayed:
• Contamination
• Electronic defective
• Sensor defective
• Venting valve defective
• Calibration error
• Temperature error
Make sure the device can no longer be operated until a qualified service techni-
cian conducts the appropriate tests and repairs.
73

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis