In the configuration menu II, tap the First nominal gas flow function field to ac-
cess the setting.
Factory setting: 0.1 l/min (0.1-1 l/min selectable)
The display switches to the First nominal gas flow menu.
Press the function field º or » to set the First nominal gas flow.
You now have the opportunity to
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the  function field to return to the previous menu level without sav-
ing,
3. or press (EXIT) to return to the work screen without saving.

10.2.3 Setting the Maximum Nominal Pressure

Operating mode
Factory setting
Bariatric
30 mm Hg
High Flow
30 mm Hg
Vessel Harvest
20 mm Hg
Pediatric
20 mm Hg
In the Configuration menu I, press the  to access the Configuration menu II.
The display changes to the Configuration menu II.
In the configuration menu, press the Maximum nominal pressure function field.
The display switches to the Maximum nominal pressure selection menu.
Press the function field º or » to specify the value.
You now have the opportunity to
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the  function field to return to the previous menu level without sav-
ing,
3. or press (EXIT) to return to the work screen without saving.
Range
5-30 mm Hg
5-30 mm Hg
5-20 mm Hg
5-20 mm Hg
Configuration Menu (Overview)
(2)
(1)
(3)
(2)
(1)
(3)
65