Using and Controlling the PNEUMO SURE High Flow Insufflator in High Flow Mode
EN
(1)
(2)
Set Pressure
Actual Pressure
mm Hg
mm Hg
0
15
Mode: High Flow
Change
START
Mode
Set Pressure
Actual Pressure
mm Hg
mm Hg
0
15
Mode: High Flow
Change
START
Mode
28
6.2
1. Selection of the insufflation mode can be done only while insufflation is
stopped. If High Flow operating mode is not yet set and displayed, press the
function field Change Mode for 2 seconds to get the insufflation mode over-
view.
2. The display depicts the different insufflation operating modes that can be ac-
tivated. Select the desired insufflation operating mode by pressing the High
Flow function field. *Inactive operating modes are displayed in gray.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the Configuration
Set Flow
menu (see chapter 10.1 Configuration menu I, page 60).
l/min
3
0.0
Liter
Actual
0
l/min
6.3
The nominal pressure can be set during insufflation or while insufflation is
stopped. Values may range from 1 to max. 30 mm Hg or the value set in the Con-
Set Flow
l/min
figuration menu.
3
Increasing/decreasing nominal pressure:
0.0
Liter
Briefly press the º or » function field to increase or decrease the pressure. Keep-
ing the º or » function field depressed longer than 1.5 seconds activates scrolling
Actual
0
l/min
in increments of 1.
Safety limit:
When increasing the nominal pressure, the status line of the display depicts Safe-
ty limit when reaching a value of 15 mm Hg. The nominal pressure 15 mm Hg is
a threshold value. This is where the recommended range for the intra-abdominal
pressure ends. Pressing the nominal pressure ºfunction field again does not in-
crease the pressure any further. Release the function field for 2 seconds. Now you
can set a value up to 30 mm Hg.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
6.4
The nominal flow can be set during insufflation or while insufflation is stopped.
1. Briefly press the º or » function field to increase or decrease the flow. The
nominal flow can be increased from 1 to max. 40 l/min.
• Briefly press the corresponding field to set values in increments of 1.

Selecting High Flow Insufflation Mode

Presetting Nominal Pressure in High Flow Operating Mode

Presetting Nominal Flow in High Flow Operating Mode