8. Serial communication

8. Serial communication
On the front of the calibrator there is a 9-pole socket
connected to the temperature controller, which enables the
calibrator to be completely controlled by a PC (see figure) by
the RS-232 input.
EN
Pin assignment, front view
The external PC must comply with the IEC 950
standard.
After switching the calibrator on and connecting the serial
interface, wait until the system has booted up.
To activate the serial communication, press the [E] button
on the display.
General characteristics
Baud rate
9600
N. Bit
8
Parity
No
Stop bit
1
The communication operates in half-duplex format which
means that transmission and reception cannot happen
simultaneously.
The controller replies only after receiving a command; it
never replies by itself.
The command and reply are ASCII character strings, as
detailed below. The communication program will be able
to convert ASCII to decimal to extract numeric values. The
default address is 1.

8.1 List of the variables and parameters

Variable
Designation
0
Set point
1
Ramp
2
Set point 2
3
Gradient
5
Proportional band
6
Integral time in seconds
22
Rx
Tx
1
2
4
3
5
6
7
8
9
Parameters
Adjustable
Readable
x ... 9999
1 = On
1 = On
0 = Off
0 = Off
x ... 9999
x ... 9999
0 ... 100 %
xxx
Variable
Designation
7
Derivative time in
seconds
10
Units
13
Cod1 (access key)
2 = preset
14
Baud rate
9600 (preset)
15
Address
16
Serial number
19
Min. set point
20
Cod2
3 = preset
21
Delay
22
Switch-on temperature
23
Switch-off temperature
Firmware version
24
27
Internal sensor type
28
Stability range
29
Stability symbol
31
Alarm setting
33
Offset of ambient
temperature
Each commands string is ASCII character succession.
First is $ character; the next must indicate the instrument
address (default 1) and then is the command (4 characters).
Possibility
RVAR
= Data reading
WVAR
= Data writing
The final part of string depends on the type of command. The
character <cr> concludes the sequence.
WIKA operating instructions, model CTD4000
Parameters
Adjustable
Readable
xxx
0 = °C
0 = °C
1 = °F
1 = °F
-
2400
4800
9600
19200
-
-
3
ON
OFF
-
-
-
-
0 = Pt100
2 = type K
-
-
0 = no
1 = yes
-
-