7.2

Preparation of the Trocar Cannula

Essential requirements
Avoid long wait times until preparation due to risk of drying and corrosion. Maximum time between use and
preparation: 2 hours. The Trocar Cannula is designed for 500 sterilization cycles.
Disassembly
Before decontamination disassemble the Trocar Cannula as described in section 6.4 "Disassembling Trocar
Cannula". Remove seal cap, internal seal and O-ring.
Manual cleaning
Rinse and clean all surfaces and lumina thoroughly with cleaning solution and suitable brushes.
Place Trocar Cannula in an ultrasonic bath with cleaning solution, if necessary.
Rinse all components of the Trocar Cannula thoroughly with cleaning gun and deionized water.
Machine cleaning and disinfection
Connect all components of the Trocar to the cleaning connections of the machine, if necessary.
Use a fine-mesh basket for small parts, such as seals and O-rings.
Run a standard machine cleaning and thermal disinfection cycle.
Autoclaving
Autoclave in vacuum autoclave (class B or S as per EN 13060) at 135°C for at least five minutes*. Do not ex-
ceed the maximum load of the sterilizer when sterilizing multiple instruments in one sterilization cycle. Au-
toclaving without vacuum must be followed by a drying phase. Allow the Trocar Cannula to dry in a bag for
at least one hour at room temperature with the paper side up.
* The temperature holding times conform to the national guidelines and standards with the maximum holding time 25 minutes.
Assembly
Make sure that all parts of the Trocar Cannula are completely dry after preparation.
The flap in the valve body must be open when screwing the spacer with internal seal to the valve body.
Assembly the Trocar Cannula in reverse order of disassembly (see chapter 6.4 "Disassembling Trocar
Cannula").