TCM 3000 BL - Morcellator
Manual disinfection
For manual disinfection, spray the electric motor, handpiece holder, motor cap with the plug and cable un-
derneath it, and the handpiece holder with isopropyl-based disinfectant spray.
Drying
If a drying programme is not provided by the washer-disinfector, the electric motor must be dried in a dry-
ing cabinet at 60°C.
Inspection, maintenance and assembly
Perform a visual inspection to check for damage, corrosion and wear and tear.
Screw the motor to spray attachment (Ref. 1974) and spray with Nou-Clean spray for about three sec-
onds and briefly wipe with a moist cloth (see instructions on spray can).
After spraying, attach the handpiece holder and screw the cable with motor cap onto the electric motor.
Packaging
Storage of single unit: Pack the electric motor in individual packaging for sterile items. The bag must be
large enough to ensure that the seal is not subject to strain. When using sterilization container, observe
the instructions of the manufacturer.
Storage of set: Sort electric motors on trays intended for this purpose or place them on all-purpose
sterilization trays.
Sterilization
Autoclave in the vacuum autoclave (Class B or S as per EN 13060) at 135°C for at least 5 minutes*.
When sterilizing several instruments during one sterilization cycle, do not exceed the maximum steriliz-
er load.
A drying cycle must be added in the case of autoclaves without a post-vacuum function. Allow the elec-
tric motor to dry in the bag for at least 1 hour at room temperature with the paper side facing upwards.
* Temperature exposure times are based on country-specific guidelines and standards. Maximum expo-
sure time: 25 minutes.
Storage
If the sterilized electric motor is not used immediately after sterilization, the material packaging must be la-
beled with the sterilization date. It is advisable to add a sterilization indicator.
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
EN
33