Herunterladen Inhalt Inhalt Diese Seite drucken

Nouvag TCM3000BL Bedienungsanleitung Seite 60

Morcellator-set
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
5.
Replacing seals
Remove the seal unit from the gear unit by rotating it
slightly.
Unscrew the membrane screw from the seal unit.
Select a suitable seal.
Keep a roof shaped seal and a membrane seal ready.
6. Position the edge of the roof shaped seal (1) to point to the
inside contact (2) of the seal holder (3) and press the roof
shaped seal in.
Position the edge of the membrane seal (4) to point to the
slot (5) of the membrane screw (6) and press the membrane
seal into the slot.
7.
Screw the membrane screw (6) into the thread of the seal
holder (3) to the stop and connect the seal unit to the gear
unit.
8. Inserting Obturator
Insert the Obturator through the seal unit, gear unit and
Trocar Cannula.
9. Attaching handle
Lift and hold the release lever (1) of the handle (2).
Push the handle (2) over the motor connection of the
gear unit.
Release the handle release lever. The handle release lev-
er must lock in position.
16
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
TCM 3000 BL - Morcellator
6
1
3
5
2
1
4
2

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis