Setting The Output Of The Irrigation Pump; Operation With Vario-Pedal - Nouvag MD 10 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
NOUVAG AG
2)
Press keys "UP" or "DOWN" to select the desired speed.
With the keys "SAW" (for Nouvag-saws) and "MUCO" (for Nouvag-Mucotome) the corresponding fixed
speed and torque values can be adjusted.
The reduction or multiplication on the keyboard panel has to be according to the
reduction or multiplication of the contra angle, otherwise the device will indicate
an incorrect revolution.

5.4 Setting the output of the irrigation pump

The output of the irrigation pump can be selected from 1 to 10 levels.
1)
Press key "PUMP SET" and in the display appears for example, pump selection [P
2)
Press keys "UP" and "DOWN" to select the desired pump speed.
3)
Press key "PUMP SET" or wait 3 seconds until the revolutions appear automatically in the display.

5.5 Operation with vario-pedal

Key "PUMP":
To turn the pump on and off (see light "PUMP SET")
Key "FORW / REV":
To adjust the direction of rotation of the motor (see light "REVERSE MODE")
Foot plate:
Push foot plate down to start pump action and to vary the motor speed.
Foot plate ...
... not pressed
... lightly pressed
... fully pressed
Caution: for safety use the unit with foot control only!
No. 31820 22/07
Motor:
Motor off
Motor runs slowly
Motor speed max.(as indicated
by setting on control unit)
6].
Pump:
Pump off
Pump on, if "PUMP" is pressed
(speed as indicated by setting on
control unit)
Pump on, if "PUMP" is pressed
(speed as indicated by setting on
control unit)
MD 10