Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
TCM 3000 BL - Morcellator
6.3

After use

1.
Switching off
Switch off the power switch on the back of the device ("O" position).
2.
Removing protective tube (when using Morcellator with protective tube)
Rotate protective tube to NO CUT position.
Close the proximal opening of the gear unit with the thumb to prevent gas loss.
Hold the protective tube and carefully disconnect gear unit with cutting tube and remove from patient.
The protective tube can also be removed from the percutaneous access.
If applicable, remove any other instruments.
3.
Removing cutting tube (when using Morcellator with Trocar Cannula)
Close the proximal opening of the gear unit with the thumb to prevent gas loss.
Hold the Trocar Cannula and carefully disconnect gear unit with cutting tube from the Trocar Cannula and
remove from patient. The valve flap closes automatically as soon as the cutting tube is pulled out.
The Trocar Cannula can remain as access throughout the remaining procedure.
4. Disconnecting gear unit connections
Pull the motor coupling from the gear unit's
motor connection.
Pull the motor device connection plug from the
motor connection of the control unit.
Pull the motor out of the handle.
Lift and hold the release lever of the handle.
Pull the handle from the gear unit's motor con-
nection.
5.
Disconnecting pedal connections
Pull the pedal device connection plug from the
pedal connection of the control unit.
6. Preparing for preparation at the location of use
Take the product to the preparation area imme-
diately.
Prepare the product as specified in this manual
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
EN
27

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis