TCM 3000 BL - Morcellator
6. Protecting tissue
The nose on the Trocar Cannula, which partially
covers the sharp edge of the blade, must protect
tissue that must not be injured.
Rotate the Morcellator to the desired position for
this.
7.
Removing tissue
Grip the required tissue with the gripper.
The handle lock is set.
Start the Morcellator set motor.
Pull the tissue towards the rotating blade with
the gripper.
Pull the gripper out of the patient through the
gear unit. The retracted mouth of the gripper
must not touch the blade.
Repeat the process until the morcellated tissue
has been completely removed.
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
Protected area
Unprotected
cutting area
EN
25