7.9
7.9.1
7.9.2
7.9.3
7.9.4
22
Downloaded from
www.Manualslib.com
For safety reasons, the unit can only be operated by pedal.
The speed of the following handpieces is fixed and cannot be changed.
● Micro saws
● Tattooing Handpiece (4 steps)

Functional check

Prior to HighSurg 30 startup or use of accessory equipment, the user must always ensure that each
individual component is in good working order, free from defects, and is clean, sterile and operational. All
inscriptions on the device and its accessories must be readable and there must be no loose parts in the
device. Once the device is switched on, depending on configuration the most recent settings entered
appear on the display and the green LED for motor 1 lights up.

Electronic motor

The performance of the electronic motor is checked without a handpiece or contra angle. Nevertheless, the
1:1 handpiece must be activated via the «Handpiece» key to check the maximum speed. Use the "Speed"
selection keys to set the speed of the electronic motor to 50,000 rpm. Press the pedal treadplate; the
electronic motor starts and accelerates to up to 50,000 rpm. When the treadplate is released, the electronic
motor slows down again.
For motor types with a permitted speed of 80,000 rpm, the maximum speed must be set accordingly to
80,000 rpm and for the Motor 16 with integrated Handpiece to 16,000 rpm.
The motor ventilation slots must be kept clear in order to prevent the motor temperature from
becoming excessive.

Pump

Press the key
on the pedal briefly; the peristaltic pump is switched on, which is shown on the display by
the symbol of a drop. Press the pedal treadplate; the peristaltic pump and the electronic motor start up.
Water sprays from the irrigation needle on the contra angle.
In the configuration menu, parameter Level 2 (Pump switch mode on/off) the pump can be set to work
permanently without pressing the treadplate. Therefore only a press on the footswitch
pump runs with the presetted value. Pressing the footswitch again stops the pump. In this case the
pressing of the pedal has no effect.

Rotational direction of the electronic motor

Press the key
on the pedal briefly; the rotational direction of the electronic motor changes. Press the
pedal treadplate; the electronic motor rotates to the left and a continuous tone is emitted. Release the
treadplate; the electronic motor ceases to operate and the tone is no longer heard. By pressing the motor
key again, the rotational direction is switched back to right rotation, which is shown on the display by the
symbol of a changing arrow.

Program

The required program can be set by repeatedly pressing the key
manuals search engine
● Dermatome
on the pedal.