SET pressure mode
The vacuum is controlled and maintained at the set pressure in the SET pressure mode.
Menu display
The following values can be changed in the menu view:
Heating bath
Rotation
Vacuum
Max. power pump
Hysteresis
Timer
Selecting the entry by turning the Select Knob Hei-GUIDE (6)
1. Call up the selected entry to be changed by pressing the Select Knob Hei-GUIDE (6)
(entry will be highlighted in orange).
2. Set the value by using the Select Knob Hei-GUIDE (6).
3. Pressing it again confi rms the change. Softkey „Esc" (1) terminates the change of the
currently processed entry.
4. The new settings are accepted by the „accept" entry. It is changed to the actual value
display.
All changes in the menu view are discarded with softkey „Esc" (1).
The parameters are stored under a name to be selected by using softkey „Store
Favorite" (1).
Actual value display
✓ The evaporator is in the SET pressure mode
Rotation speed and heating bath can be changed as described in chapter "Operation",
section "Adjust fl ask rotation" and section "Setting bath temperature".
Downloaded from
www.Manualslib.com
Operation
Setting of the bath temperature. Setting range is between
20 °C and maximum heating bath temperature (MAX TEMP is
adjustable in Setup) with a highest range of 210 °C.
Setting of rotation speed of the distillation fl ask. Setting
range is between 10 and 280 rpm.
Setting of the absolute pressure in the rotary evaporating
system. Setting range is between 1 and safety pressure
(SAFETY PRESSURE is adjustable in Setup) with a high limit
of 1200 mbar.
Vacuum pump (Vario models) throughput % to maximize
pump power. Setting range is between 10 and 100 %. Reduc-
tion of the throughput leads to higher solvent recovery rates.
The value of the hysteresis indicates how much the pressure
may rise (in mbar) before the vacuum valve is opened.
Setting range is between 1 and 50 mbar (only by using valve
controlled pumps Rotavac valve control, Rotavac valve tec, or
house vacuum source).
Setting of the timer to shut down the process. Setting range
is between 1 and 1440 minutes. For continuous operation
choose „off" for Timer setting. The heating bath at the end
of a timed run can be set to turn off (Parameter found in
Setup).
The setting range of the menu entry „vacuum" can be set
between 1 and the safety pressure in [mbar], however, not higher
than 1200 mbar.