Adjust fl ask rotation
By increasing fl ask rotation you can increase your evaporation rate. A shorter evaporation
time may prevent thermal damage to your sample.
Depending on fl ask size and heating bath temperature different rotation speeds may be
suitable.
Speed range and setting
Rotation speed can be set from 10 rpm - 280 rpm.
➜ Test your selected fl uid
▪
with slowly increasing temperatures
▪
with different evaporating fl asks
▪
with increasing rotation speeds
➜ Note these results and make them accessible for all individuals.
Setting the rotation speed
(actual value displays of all modes)
✓ The unit is switched on.
✓ The rotation is switched off (pushbutton Rotation start/stop).
✓ Actual value display
Downloaded from
www.Manualslib.com
Operation
Warning / Avertissement:
Risk of scalding and slipping!/
Risque de brûlures et de glissade !
At high fl ask rotation speeds bath fl uid may splash or spill over and
result in a slippery fl oor and/or severe scalding./
La vitesse de rotation rapide du ballon peut provoquer des éclabous-
sures ou des débordements rendant le sol glissant et/ou entraînant des
brûlures graves.
If you are operating the unit without a safety hood: /
En cas d'exploitation de l'unité sans capot de protection :
▪ Run a test with your heating bath fl uid temperature and rotation
speed and document any size of splashing./
Pratiquer un test avec la température et la vitesse de rotation du bain
chauffant et noter le volume des éclaboussures.
▪ Ensure that this documentation is communicated and accessible to all
individuals./
S'assurer que ces notes sont communiquées et mises à disposition de
tous les individus.
▪ Wear safety clothes if you are working in near proximity to the unit./
Porter des vêtements de protection lors du travail à proximité de
l'unité.
▪ Pay attention to any amount of heating bath fl uid on the fl oor./
Faire attention à toute quantité de liquide de bain chauffant répandu
sur le sol.
Depending on the oil and age of oil you are utilizing the heating bath
temperature may change the viscosity and the volume of the bath fl uid.
Keep records of the exact description/manufacturer information and
period of use and repeat the documentation in the determined time
interval.
manuals search engine
Operation
20