Connecting to a PC
The vacuum box can be connected to a PC via USB.
1. Connect the vacuum box to a PC using an USB cable.
2. Configure Hyper Terminal on the PC.
3. Interface settings:
▪ Bits per second: 115200
▪ Data bits: 8
▪ Parity: none
▪ Stop bits: 1
▪ Stream control: none
4. The protocol values are displayed in Hyper Terminal:***
(only available, if rotation is switched on)
▪ For example: 20;31.4;25.3;976
▪ Rotation = „20"
▪ HB-Temperature = „31.4"
▪ Ext. Sensor-Temperature = „25.3"
▪ Vacuum = „976"

Setup


Online Help
Safety pressure
Heating Bath Fluid
Max. Temp. Heating
Bath
Language
Bath Power at End
Brightness Display
Calibrate p-sensor
Error history
Reboot setup
Scale time axis
Downloaded from
www.Manualslib.com
Operation
Shows Online-Help instead of the actual values in the menu
SETpressure, AUTOeasy, AUTOaccurate, Gradient, Favorites and
Setup.
If the system reaches the setting of the Safety Pressure point,
the vacuum control and flask rotation is stopped, the evaporator
is vented and the motor lift is raised (motor lift only). Setting
range is between 900 and 1200 mbar.
Allows switching between different heating bath fluid and sets a
limit to the maximum heating bath temperature for the fluid for
operator safety.
Sets the limit of the maximum heating bath temperature which
can be set. The chosen bath fluid has an influence on the
temperature range. Maximum set range between 20 and 210 °C.
Sets language of the interface.
Sets the state of the bath after the process has ended. When
setting is „off" the bath is powered down after the process is
terminated.
Changes brightness of the display.
Opens sub-menu for calibration of the pressure sensor.
Shows the errors and history of occurrence for service data.
All settings of the setup are put into the state of factory setting
when Reboot engaged.
Sets range of the visible time frame in the graphical „display
actual values". Set range is between 1 and 120 minutes.
manuals search engine