DIRECTIONS FOR USE

Use Aseptic Technique
FILLING THE EASYPUMP
Figure 2 (see page 4)
1. Close clamp.
2. Un-cap the fill port.
3. Attach filled syringe to fill port. Invert pump
as shown. Grasp syringe with both hands.
Push down on plunger continuously until
volume is dispensed. Do not handle pump
while filling, as the syringe tip may break.
Repeat as necessary. Note: Filling Extension
Sets are provided with larger pumps (see
product insert).
4. Do not fill less than minimum or exceed
maximum fill volume. See Easypump C-bloc
RA Pump Insert for model specific information
on fill volumes.
5. Replace fill port cap. Label with the appropriate
pharmaceutical and patient information.
PRIMING THE ADMINISTRATION SET
Refer to Easypump C-bloc RA Pump Insert for
model specific information for priming the
administration set, starting the infusion, and
Flow Controller information.
END OF INFUSION
Figure 3 (see page 4)
Infusion is complete when pump is no longer
inflated. Dispose of pump according to your
institution's protocol.
CAUTIONS
1. The nominal infusion rate and fill volume for
each pump is labeled on the fill port.
2. Actual infusion times may vary due to:
• Filling the pump less than nominal results in
faster flow rate.
• Filling the pump greater than nominal results
in slower flow rate.
• Viscosity and/or drug concentration.
• Positioning the pump
above (increases flow rate) or below
(decreases flow rate) the catheter site.
• Temperature: Refer to Easypump C-bloc RA
Pump Insert for model specific information on
temperature.
3.This product uses DEHP plasticized PVC.
Certain solutions may be incompatible with
the PVC material used in the administration
set. Consult the drug package insert and
other available sources of information for
a more thorough understanding of possible
incompatibility problems.
4.If refrigerated, allow pump to warm to room
temperature before using.
5. Start delivery within 8 hours of filling.
Storage of a filled Easypump for more than 8
hours prior to starting infusion may result in a
slower flow rate.
6. Avoid contact of cleansing agents (like soap
and alcohol) with the filter because leakage
may occur from the air eliminating vent.
7.Roll tubing between fingers to promote flow if
clamped for extended time.
NOTE
Latex is not in fluid pathway or in contact with
human. Technical Bulletin available upon request.

STORAGE CONDITIONS

Store protected from sunlight,
40°C
10°–40° C, 10–90% relative
10°C
humidity.
Rx only =
CAUTION: Federal (U.S.A.) law
restricts this device to sell by or on the order of
a physician.
U.S. Patents: 5,080,652; 5,284,481; U.S. and
Foreign Patents Pending.
® C-bloc and Easypump are registered trademarks of
I-Flow Corporation.