6.1.4 - CALIBRATION menu

The calibration procedure is recommended to
compensate for any possible drift of the tool torque or
after each change of tool element.
6.1.4.1 - SPINDLE MANu menu
This menu is used to calculate and to apply a torque
correction coefficient to the torque value of the selected
cycle.
The torque transducer inserted in line with the tool can
be connected to any measuring unit in the Desoutter
range.
Run a tightening cycle 5 times and manually enter the
values read on the standard instrument.
● The Reset value key resets the readings.
● The Reset coeff. key displays coefficient 1 by default.
Depending on the option selected (K Torque/spindle or K
Torque/cycle) in the "STATION menu", page 26, the Torque
correction coefficient is saved:
● Either in the tool memory.
● Or in the controller.
The torque and angle reports MuST be
correct to allow the procedure to be
processed in normal conditions.
04/2010
Original instructions
6.1.4.2 - SPINDLE AuTO menu
The target of this operation is to recalibrate completely
the tool, for example after a motor change or a tool
electronic change. This needs skilled operators.
Equipment required:
● A torque measuring unit DELTA connected to the
CVIC controller via a serial cable.
● The tool to be calibrated with a transducer and its
cable.
Program the measuring unit by pressing
display [standard] in the summary line, then
display [Calib CVIC].
Select the type of transducer to be used by pressing
then
If the measuring unit is not correctly connected or
programmed, an error message [Wait. for conn] is
displayed on the screen.
Follow the instructions displayed on the CVIC screen.
10 tests can be run and they are performed at various
increasing speeds.
Run one test after another.
Press

6.1.4.3 - Calibration service

For full certified calibration, to cover your quality systems
needs please consult your local Desoutter Customer
Service Center that are fully prepared to support you,
either at your site or in one of our workshops.
Being the equipment manufacture we are prepared not
only to provide the calibration service and certification,
but also to adjust your equipment for its fullest
performance.
Our labs can provide you with either a local traceability
chain to National standards or a International level,
through ISO 17025 certified labs

6.2 - SERVICE menu

See "Start up", page 10.
The tool will be tuned over its entire
operating torque range.
/
.
The tightening is performed up to the MAX
torque.
to validate the writing in the tool memory.