6.6

Using the product

During use, check the product for damage, including all insula-
tions.
Press the yellow button to activate cutting mode. Press the blue
button to activate coagulation mode. Make sure that the electro-
surgical unit responds correctly.
Using the light gray ReMode key, you can, in the case of a pro-
gram with two setting options a and b toggle between both set-
ting options.
In order to move the electrode longitudinally, turn the rotary
wheel, or alternatively, move the shaft tube back and forth.
Spatula electrodes can be turned freely in any position. To do so,
turn the shaft tube.
In the case of laparoscopic use: retract the electrode completely
when introducing the applicator or removing it from the trocar.
You must feel the rotary wheel clicking into place. Only introduce
the applicator into the trocar with visual control.
Art. no. 20132-256: This applicator has a flexible shaft tube
(WARNING! Only bend by hand, do not use any tools) and only
one blue key. The blue button activates coagulation mode. Make
sure that the electrosurgical unit responds correctly.
7

Disposal

Dispose of the product, packaging material and accessories (if
available) in accordance with the national guidelines and regu-
lation applicable in each case.