TEST CORRECT IF: P3 - P2 < 13 mbar
5. OPERATION
5.1. AVAILABLE STERILISATION PROGRAMMES
NOTE: The length of the cycle and the amount of water used depend on the load and the initial temperature level of the autoclave. The following values have been calculated for
the maximum load and with the autoclave at room temperature.
TYPE OF CYCLE
MATERIAL TO STERILISE
QUAZ 17 l
QUAZ 24 l
- Please follow the manufacturer's recommendations
- For each programme, take into account the established maximum mass of the objects that are going to be sterilised, checked and approved by the manufacturer.
IMPORTANT:
- The BOWIE & DICK test checks the penetration of steam in the sterilisation chamber.
- When the cycle has finished the QUAZ autoclave saves and displays on the screen if
the cycle is correct or not, ANALYSING ONLY THE STERILISATION PARAMETERS THAT THE
MACHINE HAS REACHED. The user should check the result of the cycle according to the
colour of the strip, using the package manufacturer's recommendations as a reference
for the test.
'B' CYCLE
'B' CYCLE
'B' CYCLE