Upgrading the EMP KKS E60 to the EMP KKS E60PLUS
8
Upgrading the EMP KKS E60 to the EMP KKS E60
Personnel:
To upgrade the EMP KKS E60 to the EMP KKS E60
functions, obtain the dongle box or the MicroFlow
The dongle box is a slot extension, as well as a means of connecting an oval gear meter
for measuring the metering rate.
If the dongle box is fitted and is activated by switching the pump off and then on again,
+
"E60
" appears on the bottom left of the pump's display unit.
If an OGM
PLUS
by switching the pump off and then on again, "E60
and the oval gear meter function is automatically enabled under Configuration in the
Pump menu.
In addition to the slot expansion, the MicroFlow
sensor to be analysed via conductivity measurements.
When the MicroFlow box is connected and is activated by switching the pump off and then
on again, "E60
MicroFlow function is automatically enabled in the appropriate sub-menu under
Configuration in the Pump menu.
If the pump software recognises that there is a dongle box, OGM
connected, and the communication is then interrupted between the pump and the
peripheral unit (through the removal of the unit, for example), an error message is
Ä Chapter 12.2 'Alarm messages (display)' on page 113).
displayed (
To prevent this error message appearing repeatedly, the communication (connection)
must be restored, or the analysis must be disabled in the pump software
Ä Chapter 12.2.1 'Disabling dongle box, OGM
(
on page 114).
The only outward difference between a dongle box and a MicroFlow box is a sticker with
the appropriate item number:
dongle box Art. No. 248606, EBS No. 10016094,
MicroFlow box Art. No. 248611, EBS No. on request.
417101442 Rev. 12-02.2019
Qualified electrician
n
n
Mechanic
n
Service personnel
n
Specialist
(oval gear meter) is connected to slot V on the dongle box and is activated
+
microflow" appears on the bottom left of the pump's display unit, and the
PLUS
and thus access the additional
PLUS
from our range of accessories.
++
" appears on the pump's display unit,
PLUS
enables a special flow rate monitoring
PLUS
, MicroFlow
48
PLUS
PLUS
or MicroFlow box
PLUS
in the pump software '