16
Technical data
DYSSUCTION
SUCTION DEVICE M
®
Unit suction capacity*
(measured at tube connector)
Vacuum
Accuracy of the pressure
display
Mains voltage mains unit
max. load current
permissible input voltage
Power consumption at 12 V
Battery, rechargeable
Charging time for flat battery
Energy of the battery pack
Run time in battery operation
Run time in mains operation
Applied part
Protection class in accordance
with IEC 60601-1
IP degree of protection
Sound emission
Ambient conditions
Device dimensions
(H x W x D)
Page 38 of 44
REF: 67900-OS
max. 8 l/min (low flow)
* The details may vary depending on the altitude, the
atmospheric pressure and the air temperature.
Medium vacuum
-60 mbar to -300 mbar (in 10 mbar increments)
Conversion factor: 1 kPa
Target pressure > -120 mbar
Target pressure ≤ -120 mbar
In:
AC 100 – 240 V
∼
Out:
DC 12 V/2.0 A
2.0 A
12 V
33 W
min. 7.2 V; lithium ion battery
6 – 7 hours
<80 Wh
approx. 18 h, depending on the load on the motor
Permanent operation
Type BF
II
IP33
Operation:
High priority alarm:
Low priority alarm:
Transport/storage
Ambient temperature:
Relative humidity:
Air pressure:
Operation
Ambient temperature:
Relative humidity:
Air pressure:
165 mm x 220 mm x 90 mm
7.5 mmHg; 1 kPa
10 mbar
∼
∼
max. Δ 5%
max. Δ 10%
/50 – 60 Hz/580 – 320 mA
35 dB (A)
53 dB (A)
52 dB (A)
-25 °C to +60 °C
to 93%, non-condensing
700 hPa to 1060 hPa
+5 °C to +40 °C
15% to 93%, non-condensing
825 hPa to 1060 hPa