Use only clean accessories that are undamaged.
•
ADDITIONAL PRECAUTIONS FOR SAFE USE IN THE HOME ENVIRONMENT
Keep the device away from children and pets.
•
Make sure that lint and dust are removed from the device.
•
It may be difficult to read the display under unfavourable light conditions.
•
BEWARE OF DAMAGE TO THE DEVICE BY PENETRATION OF LIQUIDS
Protect the device from moisture and clean it only as described in these instructions for use.
If the interior of the device comes into direct contact with liquids or solids, it must be
inspected by Andreas Fahl Medizintechnik–Vertrieb GmbH.
BEWARE OF DAMAGE TO THE DEVICE BY INCORRECT ELECTRICAL CONNECTION
Use only the included ATM024T-W120 power supply unit (Art. no. 67912).
•
When using the power supply unit, connect the device plug of the power supply unit to
•
the suction device first and only then connect the power supply unit to the mains
power.
The power supply unit must be disconnected from the power supply in exactly the
•
reverse sequence.
BEWARE OF DAMAGE TO THE DEVICE BY IMPROPER HANDLING
Never suction flammable, corrosive or explosive liquids or gases.
•
Do not drop the device.
•
Do not use the device if obvious damage is visible on the case.
•
Make sure that the device cannot be subject to impact or fall so persons cannot be hit
•
by the falling device.
Never touch parts of non-medical devices at the same time in the area of the patient
•
and near the patient.
BEWARE OF DAMAGE TO THE DEVICE BY HEATING
Do not cover the power supply unit.
•
Do not expose the device to direct sunlight.
•
Keep the suction device, the power cord and the power supply unit away from other
•
heat sources.