Error message
"Standby"
operating mode
(Alarm after
15 minutes)
13
Decommissioning and storage
DECOMMISSIONING
1. Switch off the device after suction by pressing and holding the
2. Disconnect the power plug from the mains socket and then the power cord or power supply unit
from the device.
3. Remove the accessories as described in chapter 9.
4. Clean the accessories that can be used again as described in chapter 10.
5. Clean the surface of the device as described in chapter 10.
6. Store the device in its shipping carton until the next use.
STORAGE
Store the DYSSUCTION
approved storage conditions:
ambient temperature:
relative humidity:
air pressure:
Charge the battery of the DYSSUCTION
S before storing the device. This ensures that it can be operated at any time.
If the device will not be used for an extended period (approx. 10 months), the battery must be
recharged.
Page 36 of 44
Status
Possible cause
The therapy was not started
The device was not switched off
Suction Device M or the DYSSUCTION
®
Suction Device M or the DYSSUCTION
®
button for 3 seconds.
Suction Device S under the
®
-25 °C to +60 °C
up to 93%, non-condensing
700 hPa to 1060 hPa
Remedy
Start therapy
Switch off the device
Suction Device
®