5. Charge the battery completely before the initial use.
6. Inspect the device and test its functions (chapter 7.3).
7. Connect the disposable secretion container to the device (chapter 7.5).
7.3
Function test
Inspect the device and test its functions without a connected container before using the device for
therapy. Proceed as follows to test the functions:
1. Press the OK button for 1 to 2 seconds
device. The following start screen appears:
2. After 5 seconds the following image is displayed:
3. Start the pump by pressing the
An alarm does not appear
If the "system closed – canister full" alarm appears, the internal filter of the device is blocked
and must be replaced by an authorised service technician.
4. Then press a finger on the tube connector and restart the therapy.
"system closed – canister full"
alarm appears after no more
than 5 seconds
If the alarm does not appear after a second test, have the device checked by a service
technician.
5. Confirm the alarm with OK.
to switch on the
button and allow the device to run without a container.
"system closed – canister full" alarm
An alarm does not appear