7.4
Connection of the disposable secretion container for the DYSSUCTION
Suction Device M
Always keep a spare disposable secretion container at hand. This is essential for safe
operation!
Remove the disposable secretion container (250 ml) with stepped connector (A) from the
1.
packaging.
Push the container onto the guide rail (E) on the DYSSUCTION
2.
disposable secretion container audibly clicks onto the lock (B).
7.5
Connection of the disposable secretion container system for the
DYSSUCTION
FUNCTIONAL FAILURE DUE TO COLLAPSING "ONEWAY" SUCTION BAG
A leak in the "Bag" external container or in the cover of the "OneWay" suction bag may
allow ambient air to flow into the "Bag" external container. This may cause the "OneWay"
external bag to collapse. Make sure,
that the cover of the "OneWay" suction bag is tightly connected to the "Bag" external
container.
that all connections are tight and correctly connected.
that the "Bag" external container is not damaged and the T-piece is tight.
Always keep a spare "OneWay" suction bag at hand. This is essential for safe operation!
Page 20 of 44
Suction Device S
®
Disposable secretion container (250
A
ml) with stepped connector
B
Lock for container
C
Suction port
D
DYSSUCTION
Suction Device M
®
E
Guide rail for container
Fig. 6
Suction Device M until the
®
®