Troubleshooting and service
CONTACT OF THE DEVICE INTERIOR WITH LIQUIDS AND SOLIDS
If the interior of the device comes into direct contact with liquids or solids, it must be
inspected by Andreas Fahl Medizintechnik–Vertrieb GmbH.
Contact Andreas Fahl Medizintechnik–Vertrieb GmbH if a fault cannot be corrected with
the actions described here.
Fault
Possible cause
The device
Battery is flat
cannot be
Fault with keypad foil
switched on
Cannula or suction tube
blockage/kinked
Tube clamp closed
Secretion too viscous
DYSSUCTION
Suction not
Device M:
possible/no
Disposable secretion
secretion
container full
flow
DYSSUCTION
Device S:
"OneWay" suction bag (1
l) is full
Internal filter is blocked
Device is still in the
Einstellungen
The alarms are exclusively technically triggered alarms, because they are identified by
the monitoring of device-related variables.
All alarm messages must be confirmed with OK (except for "internal error").
Page 34 of 44
Remedy
Connect power supply unit
Please contact your service partner
Check cannula position or cuff pressure
Wash or replace cannula or suction tube
Adjust position of device
Open tube clamp
Contact your doctor or nursing care personnel
Suction
®
Replace disposable secretion container
Suction
®
Replace "OneWay" suction bag (1 l)
Please contact your service partner
Complete selection (see chapter 8.2) start therapy