Herunterladen Diese Seite drucken

Fahl DYSSUCTION Suction Device M Gebrauchsanweisung Seite 53

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
BEWARE OF INSUFFICIENT DEVICE POWER
When using the devices at an ambient air pressure below 825 hPa, the maximum vacuum
of -300 mbar cannot be reached to 100%.
However, this will not affect the safety and functionality of the device.
Overview of air pressure at 15 °C:
Air
Altitude above
pressure
mean sea level
1013 hPa
900 hPa
825 hPa
746 hPa
700 hPa
BEWARE OF RADIO INTERFERENCE
Other devices, examinations or treatments may be influenced by the device. For this
reason, other devices and examinations or treatments going on at the same time
should be carefully monitored to ensure that any interference will be detected as soon
as possible.
The device fulfils the requirements of EN 60601-1-2. The system will not be affected by the
electromagnetic radiation of other devices that conform to the same standard. The system
also does not generate electromagnetic radiation that exceeds the limit values of EN 60601-
1-2.
More information on electromagnetic immunity and electromagnetic radiation can be obtained
from Andreas Fahl Medizintechnik–Vertrieb GmbH by email to vertrieb@fahl.de or it can be
downloaded from our website at www.fahl.com.
0 m
1000 m
1700 m
2500 m
3000 m
Page 9 of 44