TROUBLESHOOTING
Issues Delivering Stimulation
Issue
Stimulation is not being
delivered during lead implant
procedure
Stimulation is not being
delivered during Stimulator
programming
Presentation
No motor or sensory responses are
observed during test stimulation
The patient does not report any
sensation of stimulation or symptom
relief
Resolution
If there are no blue bars in the stimulation bar when
stimulation is on, there is an issue with the connection.
Check the impedance using the CP.
Confirm the device cables are connected correctly. Replace
the cables if impedance is bad when the cables are correctly
connected.
If stimulation appears to be on, adjust the needle or
lead and stimulation amplitude until a motor or sensory
response is observed.
Check the lead impedance on the Patient Device screen.
If there is an impedance issue with the active electrode(s),
reprogram the device to only use electrodes with
acceptable impedance.
If there is still an issue, confirm that the CP is
communicating with the correct Stimulator. Try
disconnecting and reconnecting the CP to the Stimulator.