Controlling Stimulation Amplitude
2
3
1
8
3
7
2
5
4
6
7
0 to 12.5mA. At high electrode impedances, the delivered
stimulation amplitude may be lower than the programmed
amplitude.
6 When stimulation is on, the area under the slider bar will
turn dark gray with horizontal blue bars.
7 Stimulation may be set to increase amplitude automatically
if "Auto" is selected.
Stimulation amplitude can be adjusted in two modes: Manual
or Automatic.
Manual Amplitude Adjustment
Manual amplitude adjustment is the default mode and is
active when "Manual" ( 7 ) is seen. In the manual amplitude
adjustment mode, stimulation amplitude is changed by
moving the slider bar ( 4 ) or by pressing the up and down
arrows ( 3 ). Stimulation is turned on by pressing the
stimulation button ( 8 ), which lights up when stimulation
is on. When stimulation is on, the amplitude can only be
adjusted using the up and down arrows. Stimulation is turned
off by pressing the stimulation button ( 8 ) again.
Note: For test stimulation during lead placement and Stimulator
programming, the stimulation amplitude will start at the programmed level
and will not ramp from zero even if the Ramp feature is configured.
21 21