Viewing the Status of a Neurostimulator
4
60
5
3
Active Therapy ( 3 ): This area shows the current stimulation
settings for the therapy programmed to the Stimulator.
For a Neurostimulator, there is only one therapy program.
Additionally, the active stimulation settings can be adjusted in
the Stimulation Control module ( 4 ). This area is intended for
basic adjustments to the stimulation being delivered. It is not
intended for reprogramming.
Controls in this module include:
"Stimulation": Turn stimulation on or off. If stimulation is
turned on, it will come on at the stimulation level shown
below the "Stimulation" button.
"Stimulation Level": Increase or decrease the stimulation
amplitude. This allows control of the stimulation level as the
patient would with their patient Remote Control. Possible
levels range from 1‐7.
"Hibernate": Puts the Neurostimulator in hibernation.
Note: The CP, including this button, cannot be used to bring a device out of
Hibernate Mode. For more information see the following manual section on
Hibernate Mode.
Additional information can be accessed via the "Therapy
History" ( 5 ) and "Threshold Detail" ( 6 ) tabs on the Patient
Device screen.