Setting Up a New Neurostimulator
1
2
3
56
There are several steps to set up a new Neurostimulator.
A. A prompt will appear when a new Neurostimulator is
connected after a lead implant, asking for permission to
transfer threshold data that was recorded during the tined
lead placement procedure. A summary of the threshold
data will then be saved to the device for reference during
programming sessions. This threshold data summary can be
viewed in the "Threshold Detail" tab of the Patient Device
screen ( 1 ).
B. Input the serial number of the patient's Remote Control
in the Remote Control ID field ( 1 ). This pairs the Remote
Control to the Stimulator, allowing the Stimulator to receive
commands from the Remote Control.
Note: The Stimulator can only communicate with the Remote Control that is
entered in this field.
C. Enter the patient identifier associated with this device into
the Patient ID field ( 2 ).
D. Enter the date of the Stimulator implant procedure in the
Device Implant Date field ( 3 ).
These fields, including the Patient ID and the Remote Control
serial number can be updated at any time as needed.