Tined Lead Test Stimulation
4
5
7
30 30
To explore the response to stimulation with one electrode.
6
Select the electrode by pressing the electrode on the screen.
The top electrode on the CP screen represents the electrode
nearest to the tip of the lead. The rectangle around the
2
selected electrode will not be faded ( 1 ).
Adjust the stimulation amplitude ( 2 ). Press the stimulation
button to stimulate ( 3 ). (See manual section Controlling
Stimulation Amplitude for more information).
1
The stimulation amplitude will be displayed next to the
electrode. For each electrode, the threshold saved is the last
8
amplitude used to deliver stimulation.
Manually record the motor ( 4 ) and sensory ( 5 ) responses to
3
stimulation by pressing the appropriate buttons.
Select a new electrode to stimulate or move on to connect to a
Stimulator ( 6 ).
Note: Only stimulation amplitude is adjustable during test stimulation. The
stimulation frequency is 14 Hz, and the pulse width is 210μs. The stimulation
frequency can be changed to 5 Hz by exiting Lead Placement and going into
CP Settings.
Note: If test stimulation cannot be stopped using the stimulation button,
disconnect the cables from the CP.
Note: Press the impedance button to check the connection with the tined
lead. If the impedance for any electrode(s) is "Bad", adjust the tined lead
connection and the ground pad, then recheck the impedance. See manual
section Understanding Impedance Values for more information about the
qualitative impedance indicators.